Suggested voltages current as of Sept, 2019.

Detailed information about this instrument can be found on the baseline report (available upon request).

Reservations

Reservations can be made online by following this link.

You will need an approved USER account. Apply for one here.

To design a panel using Fluorofinder please click here.

LSR-II Configuration:

BLUE 488 LASER (20mW):

Detector

LP

Band Pass

Fluor's

Low Voltage

BLUE A

735

780/60

PE-Cy7, PE-Alexa750

500

BLUE B

685

710/50

PerCP-Cy5.5, PE-Cy5.5, PerCP, PE-Alexa700

550

BLUE C

665 660/20 PE-Cy5, PE-Alexa647, 7AAD, Fura-Low_Ca++ 550

BLUE D

600

610/20

630/30

PI, PE-Texas Red, PE-Alexa610, PE-Alexa594

500

BLUE E

550

576/26

PE

550

BLUE F

505

525/50

FITC, Alexa488, YFP, GFP, CFSE

400
BLUE G   488/10 SSC  

RED 633 LASER (20mW):

Detector

LP

Band Pass

Fluor's

Low Voltage

RED A

735

780/60

APC-Cy7, APC-Alexa750, APC-H7

450

RED B

685 710/50 APC-Cy5.5, AlexaFluor 700 350

RED C

 

660/20

APC, Alexa 647

300

VIOLET 405 LASER (40mW):

Detector

LP

Band Pass

Fluor's

Low Voltage

VIOLET A

595

610/20

BV 605

400

VIOLET B

505

525/50

Fura-High_Ca++, Pacific Orange

350

VIOLET C

 

450/50

AlexaFluor 405, DAPI, BV 421, Pacific Blue

300